Measurement PrinciplePrinciple
Ammonium molybdate spectrophotometry. In an acidic medium, orthophosphate reacts with a chromogenic agent and a reducing agent to form a colored complex. The absorbance is measured at a specific wavelength, which is proportional to the concentration of orthophosphate.
ØThe ten-way valve with a sapphire core has high wear resistance, resistance to strong acids and alkalis, organic solvents, long service life, and high stability;
ØUses a specially designed reference method, where minor displacement of the digestion cell or slight contamination does not affect the measurement value;
ØFeatures waste liquid separation function, enabling separate discharge of high-concentration reaction waste and low-concentration cleaning waste;
ØHas dynamic dilution function, automatically adjusting the dilution ratio based on the on-site set range;
ØIncludes built-in quality control functions such as spike recovery, blank testing, standard sample verification, zero point verification, span verification, parallel sample testing, and supports automatic, manual, and remote initiation of quality control;
ØCapable of automatic fault diagnosis and uploading/recording abnormal information, such as component failure, over-range alarm, exceedance alarm, reagent shortage alarm, etc.;
ØEquipped with overheating protection: automatically cuts off power to the heater if the temperature is too high;
ØAfter abnormal reset or power restoration, the instrument automatically discharges residual reactants and returns to the initial working state. Features power-on alarm function, generating an alarm message upon restart after power loss;
ØHas process log, measurement data recording, and alarm information storage functions. With data recorded hourly, it can store over 5 years of data and supports quick export via USB drive;
ØFeatures three-level management permissions.
Typical ApplicationsApplication
ØOnline monitoring of orthophosphate in drinking water and surface water;
ØOnline monitoring of orthophosphate in municipal sewage outlets, pollution source discharge points, industrial wastewater outlets, secondary sedimentation tanks, and high-efficiency sedimentation tanks.
Technical ParametersSpecification
Range | 0-10 mg/L, as P (customizable) 0-30 mg/L, as PO4 (customizable) Switchable between as P and as PO4 by operator and above permissions |
Indication Error | ≤±5% or ≤±0.005 mg/L, whichever is greater |
Detection Limit | ≤0.01 mg/L |
Quantitation Limit | ≤0.02 mg/L |
Repeatability | ≤2% |
Resolution | 0.0001 mg/L (decimal places configurable from 2 to 4 digits) |
Measurement Cycle | 23 minutes |
Display | 7-inch touch screen |
Measurement Modes | Continuous mode, periodic mode (adjustable from 1h to 30d), fixed-point mode, controlled mode, manual mode optional |
Calibration Modes | Manual calibration and automatic calibration |
Maintenance Cycle | Generally once per month |
Power Supply | 220VAC±10%, 50HZ±10% |
Output | 4-20mA output |
Communication | MODBUS RS-232, RS-485 |
Environmental Conditions | Indoor with adjustable temperature, recommended range +5 to 40°C; humidity ≤85% (non-condensing) |
Dimensions | 302*271*462mm (L*W*H) |
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 陕西天韵 |
| Spec: | TY-PO4200 |
| Stock: | |
| Origin: | China / Shaanxi / Xianshi |